Order Code LAB1909 Erythropoietin, Serum

Additional Codes

Mayo Test ID
EPO

Reporting Name

Erythropoietin (EPO), S

Useful For

An aid in distinguishing between primary and secondary polycythemia

 

Differentiating between appropriate secondary polycythemia (eg, high-altitude living, pulmonary disease, tobacco use) and inappropriate secondary polycythemia (eg, tumors)

 

Identifying candidates for erythropoietin (EPO) replacement therapy (eg, those with chronic renal failure)

 

Evaluating patients undergoing EPO replacement therapy who demonstrate an inadequate hematopoietic response

Specimen Type

Serum


Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 0.6 mL

Collection Instructions:

1. Morning collection, 7:30 a.m.-12 p.m. is preferred due to diurnal variation. For more information see Cautions.

2. Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Frozen  14 days
  Ambient  7 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross Icterus OK

Reference Values

2.6-18.5 mIU/mL

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Day(s) Performed

Monday through Friday

Method Name

Immunoenzymatic Assay

CPT Code Information

82668

Method Description

The Access erythropoietin (EPO) assay is a 2-site immunoenzymatic (sandwich) assay. A sample is added to a reaction vessel along with the paramagnetic particles coated with mouse monoclonal anti-EPO, blocking reagent and the alkaline phosphatase conjugate. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of EPO in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.(Instruction manual: Beckman Coulter Access EPO. Beckman Coulter, Inc; 2019)