Order Code URPRP Ureaplasma species, Molecular Detection, PCR, Plasma

Useful For

Rapid, sensitive, and specific identification of Ureaplasma urealyticum and Ureaplasma parvum from plasma

 

This test is not intended for medicolegal use.

Method Name

Real-Time Polymerase Chain Reaction (PCR) using LightCycler and Fluorescent Resonance Energy Transfer (FRET)

Reporting Name

Ureaplasma PCR, P

Specimen Type

Plasma EDTA


Specimen Required


The high sensitivity of amplification by polymerase chain reaction requires the specimen to be processed in an environment in which contamination of the specimen by Ureaplasma DNA is unlikely.

 

Collection Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Royal blue top (EDTA), pink top (EDTA), or sterile vial containing EDTA-derived aliquot

Submission Container/Tube: Screw-capped, sterile container

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot plasma into a plastic vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Plasma EDTA Refrigerated (preferred) 7 days
  Frozen  7 days

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Reference Values

Not applicable

Method Description

This polymerase chain reaction (PCR) method employs a target-specific detection system including primers, as well as fluorescent resonance energy transfer (FRET) hybridization probes designed for the ureC gene of Ureaplasma urealyticum and Ureaplasma parvum. The LightCycler instrument amplifies and monitors target nucleic acid sequences by fluorescence during PCR cycling. This is an automated PCR system that can rapidly detect amplified product development. The detection of amplified products is based on the FRET principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3' end is excited by an external light source, which emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, on the 5' end. The acceptor fluorophore then emits light of a different wavelength that is measured with a signal that is proportional to the amount of specific PCR product. The process is completed in a closed tube system and the melting temperature of the probes allows differentiation of U urealyticum from U parvum.(Cunningham SA, Mandrekar JN, Rosenblatt JE, Patel R: Rapid PCR detection of Mycoplasma hominis, Ureaplasma urealyticum, and Ureaplasma parvum. Int J Bacteriol 2013;2013:168742 doi: 10.1155/2013/168742)

Day(s) Performed

Monday through Friday

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

87798 x 2

87999 (if appropriate for government payers)