Sign in →

Order Code HCVQN Hepatitis C Virus (HCV) RNA Detection and Quantification by Real-Time Reverse Transcription-PCR, Serum

Additional Codes

RMC Reflex Order Code: LABHCVQN

Mayo Code: HCVQN


Ordering Guidance


For detection and quantification of hepatitis C viral RNA in serum before, during, and after antiviral therapy for chronic hepatitis C.



Shipping Instructions


1. Freeze serum immediately, and ship specimen frozen on dry ice only.

2. If shipment will be delayed for more than 24 hours, freeze serum at -20 to -80° C (up to 84 days) until shipment on dry ice.



Specimen Required


Supplies: Sarstedt 5 mL Aliquot Tube (T914)

Collection Container/Tube: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 1.5 mL

Collection Instructions: Centrifuge and aliquot serum into plastic vial within 2 hours of collection.


Useful For

Detection of acute hepatitis C virus (HCV) infection before the appearance of HCV antibodies in serum (ie, <2 months from exposure)

 

Detection and confirmation of chronic HCV infection

 

Quantification of HCV RNA in serum of patients with chronic HCV infection (HCV antibody-positive)

 

Monitoring disease progression in chronic HCV infection and response to antiviral therapy

 

Determining cure and detection of relapse after completion of antiviral therapy

Method Name

Real-Time Reverse Transcription-Polymerase Chain Reaction (RT-PCR)

Reporting Name

HCV RNA Detect/Quant, S

Specimen Type

Serum SST

Specimen Minimum Volume

0.8 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum SST Frozen (preferred) 84 days
  Refrigerated  6 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus OK

Reference Values

Undetected

Method Description

The cobas hepatitis C virus (HCV) assay is a US Food and Drug Administration-approved, in vitro nucleic acid amplification test for the quantification of HCV RNA in human serum using the cobas x800 Systems for fully automated viral nucleic acid extraction (generic silica-based capture technique) and automated amplification and detection of the viral nucleic acid sequence. This polymerase chain reaction (PCR) assay amplifies sequences within the highly conserved 5' noncoding region of the HCV genome and generates amplification products that are detected and quantified in real-time with 2 sequence-specific TaqMan probes. A non-HCV armored RNA quantitation standard (RNA QS) is introduced into each specimen during sample preparation to serve as internal control for nucleic acid extraction and PCR amplification/detection processes. Fluorescent reporter dye-labeled TaqMan probes hybridized to the complementary HCV target sequences and RNA QS sequence undergo hydrolysis during PCR amplification step to generate fluorescent signal detected in 2 different dye channels. Concentration of the HCV RNA in a patient's serum sample is determined by a ratio of the intensity of the fluorescent dye from the cleaved HCV target sequence probes and that from the RNA QS target probe detected throughout the PCR process.(Package insert: cobas HCV-Quantitative nucleic acid test for use on the cobas 5800/6800/8800 Systems. Roche Molecular Systems, Inc; Rev. 5.0, 5/2023)

Day(s) Performed

Monday through Saturday

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

87522

Forms

If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:

-Gastroenterology and Hepatology Test Request (T728)

-Microbiology Test Request (T244)

-General Test Request (T239)