Order Code HSMP Hepatosplenomegaly Panel, Plasma
Ordering Guidance
This test should not be used for monitoring of patients with confirmed diagnoses. If testing requested is for monitoring purposes, see:
CTXP / Cerebrotendinous Xanthomatosis, Plasma
GPSYP / Glucopsychosine, Plasma
OXNP / Oxysterols, Plasma
Specimen Required
Collection Container/Tube:
Preferred: Lavender top (EDTA)
Acceptable: Green top (sodium heparin, lithium heparin), yellow top (ACD B)
Submission Container/Tube: Plastic vial
Specimen Volume: 0.3 mL
Collection Instructions:
1. Centrifuge at 4° C, if possible
2. Aliquot plasma into plastic vial. Do not disturb or transfer the buffy coat layer.
3. Send frozen
Forms
If not ordering electronically, complete, print, and send a Biochemical Genetics Test Request (T798) with the specimen.
Useful For
As a component to the initial evaluation of a patient presenting with hepatosplenomegaly, using plasma specimens
This test is not useful for the identification of carriers.
This test should not be used as a monitoring tool for patients with confirmed diagnoses.
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Reporting Name
Hepatosplenomegaly Panel, PSpecimen Type
PlasmaSpecimen Minimum Volume
0.25 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma | Frozen | 65 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
CHOLESTANE-3-BETA, 5-ALPHA, 6-BETA-TRIOL
Cutoff: ≤0.070 nmol/mL
7-KETOCHOLESTEROL
Cutoff: ≤0.100 nmol/mL
LYSO-SPHINGOMYELIN
Cutoff: ≤0.100 nmol/mL
GLUCOPSYCHOSINE
Cutoff: ≤0.003 nmol/mL
7-ALPHA-HYDROXY-4-CHOLESTEN-3-ONE (7a-C4)
Cutoff: ≤0.300 nmol/mL
7-ALPHA,12-ALPHA-DIHYDROXYCHOLEST-4-en-3-ONE (7a12aC4)
Cutoff: ≤0.100 nmol/mL
Method Description
An internal standard is added to an aliquot of plasma, which is then subjected to protein precipitation. Following centrifugation, the supernatant is subjected to liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis. The MS/MS is operated in the multiple reaction monitoring positive mode to follow the precursor to product species transitions for each analyte and internal standard. The ratio of the extracted peak areas to internal standard determined by the LC-MS/MS is used to calculate the concentration of in the sample.(Unpublished Mayo method)
Day(s) Performed
Tuesday, Thursday
Performing Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
82542