Order Code LAB105 Folate, Serum

Additional Codes

Mayo Test Code: FOL

Reporting Name

Folate, S

Useful For

Investigation of suspected folate deficiency

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Patient preparation:

1. Patient should be fasting for 8 hours.

2. Do not order on patients who have recently received methotrexate or other folic acid antagonists.

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume:0.6 mL

Collection Instructions:

1. Serum gel tubes should be centrifuged within 2 hours of collection.

2. Red-top tubes should be centrifuged, and the serum aliquoted into a plastic vial within 2 hours of collection.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 7 days
  Frozen  90 days

Special Instructions

Reference Values

≥4.0 mcg/L

<4.0 mcg/L suggests folate deficiency

Day(s) Performed

Monday through Saturday

CPT Code Information

82746

Method Description

The instrument used is a Beckman Coulter DXI 800. The Access Folate assay is a competitive-binding receptor assay. A serum specimen is treated to release folate from endogenous binding proteins. After neutralization of the reaction mixture, folate-binding protein, mouse antifolate-binding protein, folic acid-alkaline phosphatase conjugate, and goat antimouse capture antibody coupled to paramagnetic particles are added to the reaction vessel. Folate in the sample competes with the folic acid-alkaline phosphatase conjugate for binding sites on a limited amount of folate-binding protein. Resulting complexes bind to the solid phase via mouse antifolate binding protein. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field, while unbound materials are washed away. The chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of folate in the sample. The amount of analyte in the sample is determined from a stored, multipoint calibration curve. The assay is standardized to the World Health Organization International Standard 03/178.(Package insert: Access Folate. Beckman Coulter Inc.; 04/2020)

Reject Due To

Gross hemolysis Reject
Gross lipemia OK

Method Name

Competitive-Binding Receptor Assay

Forms

If not ordering electronically, complete, print, and send a Benign Hematology Test Request Form (T755) with the specimen.