Order Code LABCARNS Carnitine, Serum
Additional Codes
Mayo Test Code: CARNS
Reporting Name
Carnitine, SUseful For
Evaluation of patients with a clinical suspicion of a wide range of conditions including organic acidemias, fatty acid oxidation disorders, and primary carnitine deficiency using serum specimens
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumNecessary Information
Patient's age is required.
Specimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube:
Preferred: Red top
Acceptable: Serum gel
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Frozen (preferred) | 60 days | |
Refrigerated | 21 days | ||
Ambient | 7 days |
Reference Values
|
Total carnitine (TC) |
Free carnitine (FC) |
Acylcarnitine (AC) |
AC/FC Ratio |
Age Group |
Range* |
Range* |
Range* |
Range |
≤1 day |
23-68 |
12-36 |
7-37 |
0.4-1.7 |
2-7 days |
17-41 |
10-21 |
3-24 |
0.2-1.4 |
8-31 days |
19-59 |
12-46 |
4-15 |
0.1-0.7 |
32 days-12 months |
38-68 |
27-49 |
7-19 |
0.2-0.5 |
13 months-6 years |
35-84 |
24-63 |
4-28 |
0.1-0.8 |
7-10 years |
28-83 |
22-66 |
3-32 |
0.1-0.9 |
11-17 years |
34-77 |
22-65 |
4-29 |
0.1-0.9 |
≥18 years |
34-78 |
25-54 |
5-30 |
0.1-0.8 |
Schmidt-Sommerfeld E, Werner E, Penn D: Carnitine plasma concentrations in 353 metabolically healthy children. Eur J Pediatr. 1988;147:356-360
Used with permission of European Journal of Pediatrics.
Day(s) Performed
Monday through Friday
CPT Code Information
82379
Method Description
Free and total carnitines are measured by tandem mass spectrometry (MS/MS) stable isotope dilution analysis. Hydrolysis enables measurement of total carnitine, and esterified carnitine (acylcarnitine) is calculated as the difference between the total and free carnitine. Quantification is enabled using deuterium-labeled carnitine (d3-carnitine) added as internal standard. A selected reaction monitoring experiment is performed by MS/MS. The first mass spectrometer detects carnitine and d3-carnitine precursors and transmits them to a collision cell within the mass spectrometer where they are fragmented. Specific fragments derived from the carnitine and internal standard are monitored in the second mass spectrometer.(Stevens RD, Hillman SL, Worthy S, Sanders D, Millington DS. Assay for free and total carnitine in human plasma using tandem mass spectrometry. Clin Chem. 2000;46(5):727-729; Miller MJ, Cusmano-Ozog K, Oglesbee D, Young S. ACMG Laboratory Quality Assurance Committee. Laboratory analysis of acylcarnitines, 2020 update: a technical standard of the American College of Medical Genetics and Genomics [ACMG]. Genet Med. 2021;23[2]:249-258)
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Flow Injection Analysis-Tandem Mass Spectrometry (FIA-MS/MS)
Forms
If not ordering electronically, complete, print, and send a Biochemical Genetics Test Request (T798) with the specimen.