Order Code LABCARNU Carnitine, Random, Urine
Additional Codes
Mayo Test Code: CARNU
Reporting Name
Carnitine, UUseful For
Evaluation of patients with a clinical suspicion of a wide range of conditions including organic acidemias and fatty acid oxidation disorders
Monitoring carnitine treatment
Performing Laboratory

Specimen Type
UrineSpecimen Required
Supplies: Urine Tubes, 10 mL (T068)
Container/Tube: Plastic, 10-mL urine tube
Specimen Volume: 1.5 mL
Collection Instructions:
1. Collect a random urine specimen.
2. No preservative.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Urine | Frozen (preferred) | 365 days |
Refrigerated | 72 hours |
Reference Values
FREE CARNITINE:
77-214 nmol/mg of creatinine
TOTAL CARNITINE:
180-412 nmol/mg of creatinine
RATIO:
Acylcarnitine to free carnitine: 0.7-3.4
Day(s) Performed
Tuesday
CPT Code Information
82379
Method Description
Free and total carnitines are measured by tandem mass spectrometry (MS/MS) stable isotope dilution analysis. Hydrolysis enables measurement of total carnitine, and esterified carnitine (acylcarnitine) is calculated as the difference between the total and free carnitine. Quantification is enabled using deuterium-labeled carnitine (d3-carnitine) added as internal standard. A selected reaction monitoring experiment is performed by MS/MS. The first mass spectrometer detects carnitine and d3-carnitine precursors and transmits them to a collision cell within the mass spectrometer where they are fragmented. Specific fragments derived from the carnitine and internal standard are monitored in the second mass spectrometer.(Stevens RD, Hillman SL, Worthy S, Sanders D, Millington DS. Assay for free and total carnitine in human plasma using tandem mass spectrometry. Clin Chem. 2000;46(5):727-729; Miller MJ, Cusmano-Ozog K, Oglesbee D, Young S. ACMG Laboratory Quality Assurance Committee. Laboratory analysis of acylcarnitines, 2020 update: a technical standard of the American College of Medical Genetics and Genomics [ACMG]. Genet Med. 2021;23[2]:249-258)
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Method Name
Flow Injection Analysis-Tandem Mass Spectrometry (FIA-MS/MS)
Forms
If not ordering electronically, complete, print, and send a Biochemical Genetics Test Request (T798) with the specimen.