Order Code CHBVS Chronic Hepatitis B Screen, Serum
Necessary Information
Date of collection is required.
Specimen Required
Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Serum gel (red-top tubes are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 1.2 mL
Collection Instructions:
1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).
2. Aliquot serum into plastic vial.
Useful For
Diagnosis and evaluation of patients at risk for or suspected of having chronic hepatitis B
This test is not offered as a screening or confirmatory test for blood donor specimens.
This test is not useful during the "window period" of acute hepatitis B virus infection (ie, after disappearance of hepatitis B surface antigen [HBsAg] and prior to appearance of hepatitis B surface antibody).
This test is not useful as a stand-alone prenatal screening test of HBsAg status in pregnant women.
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
HBGSN | HBs Antigen Scrn, S | Yes | Yes |
Reflex Tests
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
EAG | Hepatitis Be Ag, S | Yes | No |
HBGSC | HBs Antigen Screen Confirmation, S | No | No |
HEAB | HBe Antibody, S | Yes | No |
Testing Algorithm
If the hepatitis B surface antigen (HBsAg) result is reactive, then HBsAg confirmation testing will be performed at an additional charge. If the HBsAg confirmation result is positive, then hepatitis B e antigen (HBeAg) and hepatitis B e antibody (anti-HBe) tests will be performed at an additional charge.
The following algorithms are available:
-Hepatitis B: Testing Algorithm for Screening, Diagnosis, and Management
-HBV Infection-Monitoring Before and After Liver Transplantation
Special Instructions
Method Name
Electrochemiluminescence Immunoassay (ECLIA)
Reporting Name
Chronic Hepatitis B Screen, SSpecimen Type
Serum SSTSpecimen Minimum Volume
0.9 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Serum SST | Frozen (preferred) | 90 days |
Refrigerated | 6 days | |
Ambient | 72 hours |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Method Description
Hepatitis B Surface Antigen:
The Elecsys HBsAg (hepatitis B surface antigen) II assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. HBsAg present in the patient's sample reacts with 2 biotinylated monoclonal anti-HBs and a mixture of monoclonal anti-HBs and polyclonal anti-HBs labeled with a ruthenium complex react to form a sandwich complex. After addition of streptavidin-coated microparticles (solid phase), the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode, which induces chemiluminescent emission that is measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the reaction product in the patient's sample to the cutoff index (COI) value set from reagent lot-specific assay calibration.(Package insert: Elecsys HBsAG II. Roche Diagnostics; v3.0, 02/2022)
HBsAg Confirmation:
The Elecsys HBsAg II Auto Confirm assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. This test is based on 2 parallel measurements. For the first measurement, the sample is treated with the control pretreatment reagent (PT2) prior to immunoreaction. This measurement serves as a reference. For the second measurement the sample is treated with the confirmatory pretreatment reagent (PT1) prior to immunoreaction. During incubation with confirmatory pretreatment, unlabeled polyclonal anti-HBs are bound to the sample HBsAg and thereby block the binding sites for the labeled antibodies used in the following immunoreaction. The confirmation result (%) is automatically assessed by determining the ratio of both measurements.
During testing, the auto-diluted sample is incubated with control pretreatment and confirmatory pretreatment, followed by formation of sandwich complexes of biotinylated monoclonal anti-HBs and a mixture of monoclonal anti-HBs and polyclonal anti-HBs labeled with a ruthenium complex. After addition of streptavidin-coated microparticles (solid phase), the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode, which induces chemiluminescent emission that is measured by a photomultiplier. Results are determined by comparing the electrochemiluminescence signal generated from the reaction product to the cutoff index value set from reagent lot-specific assay calibration. The confirmation result (%) is calculated from the ratio of the COI obtained for the measurement with confirmatory pretreatment to the COI obtained for the measurement with control pretreatment.(Package insert: Elecsys HBsAg II Auto Confirm. Roche Diagnostics; v1.0, 12/2020)
Day(s) Performed
Monday through Saturday
Performing Laboratory

CPT Code Information
87340
G0499-(if appropriate)
87350 (if appropriate)
87341 (if appropriate)
86707 (if appropriate)
Forms
If not ordering electronically, complete, print, and send 1 of the following: