Order Code CMVMS Cytomegalovirus Antibody, IgM, Serum
Specimen Required
Supplies: Sarstedt Aliquot Tube 5 mL (T914)
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 0.6 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Useful For
Aiding in the diagnosis of acute infection with cytomegalovirus
Method Name
Electrochemiluminescence Immunoassay (ECLIA)
Reporting Name
CMV Ab, IgM, SSpecimen Type
SerumSpecimen Minimum Volume
See Specimen Required
Specimen Stability Information
| Specimen Type | Temperature | Time |
|---|---|---|
| Serum | Refrigerated (preferred) | 28 days |
| Frozen | 90 days | |
| Ambient | 7 days |
Reject Due To
| Gross hemolysis | Reject |
| Gross lipemia | Reject |
| Gross icterus | Reject |
| Additives (eg, biocides, antioxidants) | Reject |
| Heat inactivated | Reject |
Reference Values
Negative
Reference values apply to all ages.
Method Description
The electrochemiluminescence immunoassay for the in vitro qualitative determination of IgM antibodies to cytomegalovirus (CMV) in human serum is a micro-capture test principle. During the first incubation, 6 mcL of sample are automatically prediluted 1:20 with Diluent Universal. Biotinylated monoclonal anti-h-IgM-specific antibodies are added. The second incubation, CMV-specific recombinant antigen labeled with a ruthenium complex and streptavidin-coated microparticles are added. The complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell II M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration.(Package insert: Elecsys CMV IgM. Roche Diagnostics GmbH; 11/2022)
Day(s) Performed
Monday through Saturday
Performing Laboratory
Mayo Clinic Laboratories in Rochester
CPT Code Information
86645
LOINC Code Information
| Test ID | Test Order Name | Order LOINC Value |
|---|---|---|
| CMVMS | CMV Ab, IgM, S | 30325-5 |
| Result ID | Test Result Name | Result LOINC Value |
|---|---|---|
| MCMV | CMV Ab, IgM, S | 30325-5 |