Sign in →

Order Code DGAP Dairy and Grain Allergen Profile, Serum

Additional Codes

Mayo Test ID
DAGR

Reporting Name

Dairy and Grain Allergen Profile

Useful For

Establishing a diagnosis of an allergy to dairy and grain

 

Defining the allergen responsible for eliciting signs and symptoms

 

Identifying allergens:

-Responsible for allergic response and/or anaphylactic episode

-To confirm sensitization prior to beginning immunotherapy

-To investigate the specificity of allergic reactions to insect venom allergens, drugs, or chemical allergens

Specimen Type

Serum


Ordering Guidance


For a listing of allergens available for testing, see Allergens - Immunoglobulin E (IgE) Antibodies



Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 0.7 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.5 mL
For 1 allergen: 0.3 mL
For more than 1 allergen: (0.05 mL x number of allergens) + 0.25 mL deadspace

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 14 days
  Frozen  90 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK

Reference Values

Class 

IgE kU/L 

Interpretation 

<0.10 

Negative 

0/1 

0.10-0.34 

Borderline/equivocal 

0.35-0.69 

Equivocal 

0.70-3.49 

Positive 

3.50-17.4 

Positive 

17.5-49.9 

Strongly positive 

50.0-99.9 

Strongly positive 

≥100 

Strongly positive 

Reference values apply to all ages.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Day(s) Performed

Monday through Friday

Method Name

Fluorescence Enzyme Immunoassay (FEIA)

CPT Code Information

86003 x 5

Profile Information

Test ID Reporting Name Available Separately Always Performed
EGG Egg White, IgE Yes Yes
MILK Milk, IgE Yes Yes
WHT Wheat, IgE Yes Yes
OATS Oat, IgE Yes Yes
SOY Soybean, IgE Yes Yes

Method Description

Specific IgE from the patient's serum reacts with the allergen of interest, which is covalently coupled to an ImmunoCAP. After washing away nonspecific IgE, enzyme-labeled anti-IgE antibody is added to form a complex. After incubation, unbound anti-IgE is washed away, and the bound complex incubated with a developing agent. After stopping the reaction, the fluorescence of the eluate is measured. Fluorescence is proportional to the amount of specific IgE present in the patient's sample (ie, the higher the fluorescence value, the more IgE antibody is present).(Package insert: ImmunoCAP System Specific IgE FEIA. Phadia: Rev 06/2020)

Forms

If not ordering electronically, complete, print, and send an Allergen Test Request (T236) with the specimen.