Order Code HEAB Hepatitis B Virus e Antibody, Serum
Additional Codes
Mayo Test ID |
---|
HEAB |
Reporting Name
HBe Antibody, SUseful For
Determining the presence or absence of detectable hepatitis B virus e antibody in monitoring infection status of individuals with chronic hepatitis B
Determining infectivity of hepatitis B virus (HBV) carriers
Monitoring serologic response of chronically HBV-infected patients receiving antiviral therapy
Specimen Type
Serum SSTAdditional Testing Requirements
If ordered with HBVQN / Hepatitis B Virus (HBV) DNA Detection and Quantification by Real-Time PCR, Serum; send separate vials.
Necessary Information
Date of collection is required.
Specimen Required
Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Serum gel (red-top tubes are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 0.7 mL
Collection Instructions:
1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).
2. Aliquot serum into plastic vial.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Serum SST | Frozen (preferred) | 90 days |
Refrigerated | 6 days | |
Ambient | 72 hours |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Performing Laboratory

Day(s) Performed
Monday through Saturday
Method Name
Electrochemiluminescence Immunoassay (ECLIA)
CPT Code Information
86707
Testing Algorithm
For more information see Hepatitis B: Testing Algorithm for Screening, Diagnosis, and Management
Special Instructions
Method Description
The Elecsys Anti-HBe (hepatitis B virus e antibody) assay is based on the competitive immunoassay principle and performed using an electrochemiluminescence method on the automated cobas e 801 immunochemistry analyzer. Anti-HBe present in the patient's sample binds to the added synthetic HBe antigen (HBeAg). The remaining unbound sites on the synthetic HBeAg become occupied with the added biotinylated antibodies and ruthenium complex-labeled antibodies specific for HBeAg. The entire complex becomes bound to streptavidin-coated microparticles (solid phase) via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. After unbound substances are washed away, voltage is applied to the electrode, which induces chemiluminescent emission that is measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the reaction product to the cutoff index value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HBe. Roche Diagnostics; v1.0, 12/2021)
Forms
If not ordering electronically, complete, print, and send 1 of the following with the specimen: