Order Code HIVSP HIV Antigen and Antibody Prenatal Routine Screen, Plasma


Ordering Guidance


If the specimen is obtained from either autopsy or cadaver blood sources, order HV1CD / HIV-1 and HIV-2 Antibodies for Cadaveric or Hemolyzed Specimens, Serum which is the US Food and Drug Administration-approved assay for these specimen types.

 

New York State clients: This test should not be requested for maternal/newborn HIV screening on specimens originating in New York State, due to state regulatory requirements for expedited result reporting.



Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Lavender top (EDTA)

Submission Container/Tube: Plastic vial

Specimen Volume: 1.5 mL

Collection Instructions:

1. Centrifuge blood collection tube per manufacturer’s instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).

2. Transfer plasma into a plastic vial.


Useful For

Screening for HIV-1 and HIV-2 infection in nonsymptomatic pregnant patients

 

This test should not be used as a screening or confirmatory test for blood donor specimens.

Reflex Tests

Test ID Reporting Name Available Separately Always Performed
HVDSP HIV Ab Differentiation Prenatal, P Yes No
HPP12 HIV-1/HIV-2 RNA Detect Prenatal, P Yes No
HIQNP HIV-1 RNA Detect/Quant Prenatal, P Yes No

Testing Algorithm

This test begins with HIV-1/-2 antigen and antibody (Ab) screen by electrochemiluminescence immunoassay. If the screen result is reactive, then HIV-1/-2 antibody confirmation/differentiation test by immunochromatographic method is performed at an additional charge.

 

If the following result types are obtained from the HIV-1/-2 confirmation/differentiation test, HIV-1/HIV-2 RNA detection will be performed at an additional charge:

-Negative for both HIV-1 Ab and HIV-2 Ab

-Indeterminate for HIV-1 Ab but negative for HIV-2 Ab

-Negative for HIV-1 Ab but indeterminate for HIV-2 Ab

-Indeterminate for both HIV-1 Ab and HIV-2 Ab

-Positive for both HIV-1 Ab and HIV-2 Ab

 

If the following result types are obtained from the HIV-1/-2 confirmation/differentiation, HIV-1 RNA detection and quantification will be performed at an additional charge:

-Positive for HIV-1 Ab and negative for HIV-2 Ab

-Positive for HIV-1 Ab and indeterminate for HIV-2 Ab

Method Name

Electrochemiluminescence Immunoassay (ECLIA)

Reporting Name

HIV-1/-2 Ag and Ab Prenatal Scrn, P

Specimen Type

Plasma EDTA

Specimen Minimum Volume

1 mL

Specimen Stability Information

Specimen Type Temperature Time
Plasma EDTA Frozen (preferred) 30 days
  Refrigerated  6 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
Heat-inactivated specimen Reject

Reference Values

Negative

Method Description

The Elecsys HIV DUO assay is based on the sandwich principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. HIV-1 p24 antigen (Ag) present in patient’s sample first reacts with biotinylated monoclonal anti-p24 antibodies and ruthenylated monoclonal anti-p24 antibodies, to form a sandwich complex. In a separate reaction vessel, HIV-1 and HIV-2 antibodies (Ab) present in the same sample react with biotinylated HIV-specific recombinant antigens/peptides and ruthenylated HIV-specific recombinant antigens/peptides, to form a sandwich complex. After addition of streptavidin-coated microparticles (solid phase) to each reaction vessel, the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixtures are then each aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode which induces chemiluminescent emissions that are measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the reactions to the cutoff index (COI) values set from reagent lot-specific assay calibration for the corresponding HIV-1 p24 Ag and HIV Ab.(Package insert: Elecsys HIV DUO. Roche Diagnostics; v1.0, 12/2020)

Day(s) Performed

Monday through Saturday

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

87389

G0475 (if appropriate)

Forms

If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.