Order Code HPLC High-Performance Liquid Chromatography (HPLC) Hemoglobin Variant, Blood
Specimen Required
Only orderable as part of a profile. For more information see:
-HAEV1 / Hemolytic Anemia Evaluation, Blood
-HBEL1 / Hemoglobin Electrophoresis Evaluation, Blood
-MEV1 / Methemoglobinemia Evaluation, Blood
-REVE2 / Erythrocytosis Evaluation, Blood
-THEV1 / Thalassemia and Hemoglobinopathy Evaluation, Blood and Serum
Useful For
Providing additional information, which aids in the identification of hemoglobin variants
Method Name
Only orderable as part of a profile. For more information see:
-HAEV1 / Hemolytic Anemia Evaluation, Blood
-HBEL1 / Hemoglobin Electrophoresis Evaluation, Blood
-MEV1 / Methemoglobinemia Evaluation, Blood
-REVE2 / Erythrocytosis Evaluation, Blood
-THEV1 / Thalassemia and Hemoglobinopathy Evaluation, Blood and Serum
Cation Exchange High-Performance Liquid Chromatography (HPLC)
Reporting Name
HPLC Hb Variant, BSpecimen Type
Whole Blood EDTASpecimen Minimum Volume
1 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Whole Blood EDTA | Refrigerated | 10 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Reference Values
Only orderable as part of a profile. For more information see:
-HAEV1 / Hemolytic Anemia Evaluation, Blood
-HBEL1 / Hemoglobin Electrophoresis Evaluation, Blood
-MEV1 / Methemoglobinemia Evaluation, Blood
-REVE2 / Erythrocytosis Evaluation, Blood
-THEV1 / Thalassemia and Hemoglobinopathy Evaluation, Blood and Serum
Method Description
Hemolysate of whole blood is injected into an analysis stream passing through a cation exchange column using high-performance liquid chromatography. A pre-programmed gradient controls the elution buffer mixture that also passes through the analytical cartridge. The ionic strength of the elution buffer is raised by increasing the percentage of a second buffer. As the ionic strength of the buffer increases, the more strongly retained hemoglobins elute from the cartridge. Absorbance changes are detected by a dual-wavelength filter photometer. Changes in absorbance are displayed as a chromatogram of absorbance versus time.(Huismann TH, Schroeder WA, Brodie AN, et al. Microchromotography of hemoglobins. III. A simplified procedure for the determination of hemoglobin A2. J Lab Clin Med. 1975;86:700-702; Ou CN, Buffone GJ, Reimer GL, Alpert AJ. High-performance liquid chromatography of human hemoglobins on a new cation exchanger. J Chromatogr. 1983;266:197-205; instruction manual: Bio-Rad Variant II Beta-thalassemia Short Program Instructions for Use, L70203705. Bio-Rad Laboratories, Inc; 11/2011)
Performing Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
83021
Day(s) Performed
Monday through Friday