Order Code LAB1107 Carbon Monoxide, Whole Blood
Useful For
Verifying carbon monoxide toxicity in cases of suspected exposure.
Specimen Requirements
Special Note: Specimen must arrive within 3 hours of draw. Draw a full, dark green-top (lithium heparin) tube. Allow tube to fill completely. Do not uncap tube once specimen has been drawn. Tube should be transported directly to laboratory.
Specimen Type: Whole Blood
Collection Container/Tube: dark green-top (lithium or sodium heparin) tube
Preferred: Dark green Lithium Heparin
Specimen Stability
Specimen Type | Temperature | Stability |
Whole Blood | Ambient | 3 hours |
Refrigerated | N/A | |
Frozen | N/A |
Reference Values
- 0 - 1.5%
- Literature indicates COHb less than 3% are considered normal for urban living nonsmoking populations.
- Critical value (automatic call-back): > 15%
Day(s) Test Set Up
Monday through Sunday
Performing Laboratory
Ridgeview Medical Center Laboratory
Two Twelve Medical Center Laboratory
Methodology
Co-oximetry
Test Classification and CPT Coding
82375