Order Code LAB1114 Haptoglobin, Serum
Additional Codes
Mayo Test ID |
---|
HAPT |
Reporting Name
Haptoglobin, SUseful For
Confirmation of intravascular hemolysis
Specimen Type
SerumSpecimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume:1 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 14 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Reference Values
30-200 mg/dL
Performing Laboratory
Mayo Clinic Laboratories in RochesterDay(s) Performed
Monday through Friday
Method Name
Nephelometry
CPT Code Information
83010
Method Description
In this Siemens Nephelometer II method, the light scattered onto the antigen-antibody complexes is measured. The intensity of the measured scattered light is proportional to the amount of antigen-antibody complexes in the sample under certain conditions. If the antibody volume is kept constant, the signal behaves proportionally to the antigen volume.
A reference curve is generated by a standard with a known antigen content on which the scattered light signals of the samples can be evaluated and calculated as an antigen concentration. Antigen-antibody complexes are formed when a sample containing antigen and the corresponding antiserum are put into a cuvette. A light beam is generated with a light emitting diode (LED), which is transmitted through the cuvette. The light is scattered onto the immuno-complexes that are present. Antigen and antibody are mixed in the initial measurement, but no complex is formed yet. An antigen-antibody complex is formed in the final measurement.
The result is calculated by subtracting value of the final measurement from the initial measurement. The distribution of intensity of the scattered light depends on the ratio of the particle size of the antigen-antibody complexes to the radiated wavelength.(Instruction manual: Siemens Nephelometer II Operations Instruction Manual; Siemens, Inc; Version 2.4, 07/2019; Addendum to the Instruction Manual 2.3, 08/2017)