Sign in →

Order Code LAB1241 C1 Esterase Inhibitor Antigen, Serum

Additional Codes

Mayo Test Code: C1ES

Reporting Name

C1 Esterase Inhibitor Antigen, S

Useful For

Diagnosis of hereditary angioedema

 

Monitoring levels of C1 esterase inhibitor in response to therapy

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Patient Preparations: Fasting preferred but not required.

Collection Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Immediately after specimen collection, place the tube on wet ice.

2. Centrifuge and aliquot serum into plastic vial.

3. Freeze specimen within 30 minutes.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Frozen (preferred) 28 days
  Refrigerated  28 days
  Ambient  72 hours

Reference Values

19-37 mg/dL

Day(s) Performed

Monday through Friday

CPT Code Information

83883

Method Description

C1 esterase inhibitor complement antigen (C1ES) is measured by immunonephelometry. Antiserum to C1ES is mixed with patient serum, the light scatter resulting from the antibody interaction with C1ES is measured, and the signal is compared to standard concentrations of C1ES.(Instruction manual: Siemens Nephelometer II Operations. Siemens, Inc; Version 2.4, 07/2019; Addendum to the Instruction Manual 2.3, 08/2017)

Reject Due To

Gross hemolysis OK
Gross lipemia Reject
Gross icterus OK

Method Name

Nephelometry