Order Code LAB1248 Carnitine, Plasma
Additional Codes
Mayo Test ID |
---|
CARN |
Reporting Name
Carnitine, PUseful For
Evaluation of patients with a clinical suspicion of a wide range of conditions including organic acidemias, fatty acid oxidation disorders, and primary carnitine deficiency using plasma specimens
Specimen Type
PlasmaNecessary Information
Patient's age is required.
Specimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube:
Preferred: Green top (sodium heparin)
Acceptable: Lavender top (EDTA), green top (lithium heparin)
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions: Centrifuge and aliquot plasma into a plastic vial
Specimen Minimum Volume
0.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Plasma | Frozen (preferred) | 60 days |
Refrigerated | 21 days | |
Ambient | 7 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
|
Total carnitine (TC) |
Free carnitine (FC) |
Acylcarnitine (AC) |
AC/FC Ratio |
Age Group |
Range* |
Range* |
Range* |
Range |
≤1 day |
23-68 |
12-36 |
7-37 |
0.4-1.7 |
2-7 days |
17-41 |
10-21 |
3-24 |
0.2-1.4 |
8-31 days |
19-59 |
12-46 |
4-15 |
0.1-0.7 |
32 days-12 months |
38-68 |
27-49 |
7-19 |
0.2-0.5 |
13 months-6 years |
35-84 |
24-63 |
4-28 |
0.1-0.8 |
7-10 years |
28-83 |
22-66 |
3-32 |
0.1-0.9 |
11-17 years |
34-77 |
22-65 |
4-29 |
0.1-0.9 |
≥18 years |
34-78 |
25-54 |
5-30 |
0.1-0.8 |
*Values expressed as nmol/mL
Schmidt-Sommerfeld E, Werner E, Penn D: Carnitine plasma concentrations in 353 metabolically healthy children. Eur J Pediatr. 1988;147:356-360
Used with permission of European Journal of Pediatrics.
Performing Laboratory

Day(s) Performed
Monday through Friday
Method Name
Flow Injection Analysis-Tandem Mass Spectrometry (FIA-MS/MS)
CPT Code Information
82379
Method Description
Free and total carnitines are measured by tandem mass spectrometry (MS/MS) stable isotope dilution analysis. Hydrolysis enables measurement of total carnitine, and esterified carnitine (acylcarnitine) is calculated as the difference between the total and free carnitine. Quantification is enabled using deuterium-labeled carnitine (d3-carnitine) added as internal standard. A selected reaction monitoring experiment is performed by MS/MS. The first mass spectrometer detects carnitine and d3-carnitine precursors and transmits them to a collision cell within the mass spectrometer where they are fragmented. Specific fragments derived from the carnitine and internal standard are monitored in the second mass spectrometer.(Stevens RD, Hillman SL, Worthy S, Sanders D, Millington DS. Assay for free and total carnitine in human plasma using tandem mass spectrometry. Clin Chem. 2000;46(5):727-729; Miller MJ, Cusmano-Ozog K, Oglesbee D, Young S. ACMG Laboratory Quality Assurance Committee. Laboratory analysis of acylcarnitines, 2020 update: a technical standard of the American College of Medical Genetics and Genomics [ACMG]. Genet Med. 2021;23[2]:249-258)
Forms
If not ordering electronically, complete, print, and send a Biochemical Genetics Test Request (T798) with the specimen.