Order Code LAB1285 Glomerular Basement Membrane Antibodies, IgG, Serum
Additional Codes
Mayo Test Code: GBM
Reporting Name
Glomerular Basement Membrane IgG AbUseful For
Evaluating patients with clinical features of anti-glomerular basement disease, including rapidly progressive glomerulonephritis or pulmonary hemorrhage
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumOrdering Guidance
If patient is being evaluated for autoimmune skin disease, order CIFS / Cutaneous Immunofluorescence Antibodies (IgG), Serum for evaluation of anti-intercellular substance (ICS) and antibasement membrane zone (BMZ) antibodies.
Specimen Required
Supplies: Sarstedt Aliquot Tube 5 mL (T914)
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.35 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 21 days | |
Frozen | 21 days |
Reference Values
<1.0 U (negative)
≥1.0 U (positive)
Reference values apply to all ages.
Day(s) Performed
Monday through Friday
CPT Code Information
83516
Method Description
Glomerular basement membrane (GBM) antigen is covalently coupled to polystyrene microspheres that are impregnated with fluorescent dyes to create a unique fluorescent signature. GBM antibodies, if present in diluted serum, bind to the GBM antigen on the microspheres. The microspheres are washed to remove extraneous serum proteins. Phycoerythrin (PE)-conjugated antihuman-IgG antibody is then added to detect IgG anti-GBM bound to the microspheres. The microspheres are washed to remove unbound conjugate, and bound conjugate is detected by laser photometry. A primary laser reveals the fluorescent signature of each microsphere to distinguish it from microspheres that are labeled with other antigens. A secondary laser reveals the level of PE fluorescence associated with each microsphere. Results are calculated by comparing the median fluorescence response for GBM microspheres to a 4-point calibration curve.(Package insert: Bio-Plex 2200 Vasculitis. Bio-Rad Laboratories; 12/2018)
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | OK |
Method Name
Multiplex Flow Immunoassay
Forms
If not ordering electronically, complete, print, and send a Renal Diagnostics Test Request (T830) with the specimen.