Sign in →

Order Code LAB4568 Cytomegalovirus (CMV) Molecular Detection, PCR, Varies

Additional Codes

Mayo Test Code: CMVPV


Ordering Guidance


For plasma specimens order CMVQN / Cytomegalovirus (CMV) DNA Detection and Quantification by Real-Time PCR, Plasma.



Necessary Information


Specimen source is required.



Specimen Required


Submit only 1 of the following specimens:

 

Specimen Type: Body fluid

Sources: Spinal, pleural, peritoneal, ascites, pericardial, amniotic, or ocular

Supplies: Sarstedt Aliquot Tube, 5mL (T914)

Container/Tube:

Preferred: Sterile, screwcap, 5-mL aliquot tube

Acceptable: Sterile container

Specimen Volume: 0.5 mL

Collection Instructions: Do not centrifuge.

 

Specimen Type: Respiratory fluid

Sources: Bronchial washing, bronchoalveolar lavage, nasopharyngeal aspirate or washing, sputum, or tracheal aspirate

Supplies: Sarstedt Aliquot Tube, 5mL

Container/Tube:

Preferred: Sterile, screwcap, 5-mL aliquot tube

Acceptable: Sterile container

Specimen Volume: 1.5 mL

Collection Instructions: Do not centrifuge.

 

Specimen Type: Genital swab

Sources: Cervix, vagina, urethra, anal/rectal, or other genital sources

Supplies:

-Culturette (BBL Culture Swab) (T092)

-M4-RT (T605)

Container/Tube: Multimicrobe media (M4-RT) and ESwabs

Collection Instructions: Place swab back into multimicrobe media (M4-RT, M4, or M5)

 

Specimen Type: Swab

Sources: Dermal, eye, nasal, saliva, or throat

Supplies:

-Culturette (BBL Culture Swab) (T092)

-M4-RT (T605)

Container/Tube: Multimicrobe media (M4-RT) and ESwabs

Collection Instructions: Place swab back into multimicrobe media (M4-RT, M4, or M5)

 

Specimen Type: Tissue

Sources: Brain, colon, kidney, liver, lung, etc.

Supplies: M4-RT (T605)

Container/Tube: Sterile container containing 1 mL to 2 mL of sterile saline or multimicrobe medium (M4-RT, M4, or M5)

Specimen Volume: Entire collection

Collection Instructions: Submit only fresh tissue in multimicrobe media (M4-RT) or a sterile container with 1 to 2 mL sterile saline

 

Specimen Type: Urine

Container/Tube: Sterile container

Specimen Volume: 1 mL

Collection Instructions: Collect a random urine specimen.

 

Specimen Type: Bone marrow

Container/Tube: Lavender top (EDTA)

Specimen Volume: 0.5 mL

Collection Instructions: Send bone marrow in original tube. Do not aliquot.


Useful For

Rapid qualitative detection of cytomegalovirus (CMV) DNA

 

This test is not intended for the monitoring of CMV disease progression.

Method Name

Real-Time Polymerase Chain Reaction (PCR)/DNA Probe Hybridization

Reporting Name

Cytomegalovirus, PCR

Specimen Type

Varies

Specimen Minimum Volume

Ocular Fluid, Spinal Fluid, or Urine: 0.3 mL
Body Fluid (pleural, peritoneal, ascites, pericardial): See Specimen Required
Respiratory Specimens: 1 mL
Tissue: 2 × 2-mm biopsy

Specimen Stability Information

Specimen Type Temperature Time Special Container
Varies Refrigerated (preferred) 7 days
  Frozen  7 days

Reject Due To

 
Calcium alginate-tipped swab
Wood swab
Transport swab containing gel
Feces
Paraffin blocks
Breast milk
Heat inactivated specimens
Reject
 

Reference Values

Negative

Reference values apply to all ages.

Method Description

Viral nucleic acid is extracted by the MagNA Pure automated instrument (Roche Applied Science) from clinical specimens. Primers directed to the target Us9 gene produce a 285-base pair amplicon. The LightCycler instrument amplifies and monitors by fluorescence the development of target nucleic acid sequences after the annealing step during polymerase chain reaction (PCR) cycling. This is an automated PCR system that can rapidly detect amplicon development through stringent air-controlled temperature cycling in capillary cuvettes. The detection of amplified products is based on the fluorescence resonance energy transfer (FRET) principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3'-end is excited by an external light source and emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, at the 5'-end. The acceptor fluorophore then emits a light of a different wavelength that can be measured with a signal that is proportional to the amount of specific PCR product. Melting curve analysis is performed following PCR amplification. Starting at 45° C, the temperature in the thermal chamber is slowly raised to 80° C and the fluorescence is measured at frequent intervals. Analysis of the PCR amplification and probe melting curves is accomplished through the use of LightCycler software.(Binnicker MJ, Espy M. Comparison of six real-time PCR assays for the qualitative detection of cytomegalovirus in clinical specimens. J Clin Microbiol. 2013:51[11]:3749-3752)

Day(s) Performed

Monday through Sunday

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

87496

Forms

If not ordering electronically, complete, print, and send a Microbiology Test Request (T244) with the specimen.