Order Code LAB5036 Hepatitis B Virus Core IgM Antibody, Serum

Additional Codes

Mayo Test ID
HBIM

Reporting Name

HBc IgM Ab, S

Useful For

Diagnosis of acute hepatitis B virus (HBV) infection

 

Identifying acute HBV infection in the serologic window period when HBV surface antigen and HBV surface antibody results are negative

 

Differentiation between acute, chronic, or past HBV infections in the presence of positive hepatitis B virus core total antibodies

Specimen Type

Serum SST


Necessary Information


Date of collection is required.



Specimen Required


Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Serum gel (red-top tubes are not acceptable)

Submission Container/Tube: Plastic vial

Specimen Volume: 0.6 mL

Collection Instructions:

1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).

2. Aliquot serum into a plastic tube.


Specimen Minimum Volume

0.6 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum SST Frozen (preferred) 90 days
  Refrigerated  6 days
  Ambient  72 hours

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject

Reference Values

Negative

 

See Viral Hepatitis Serologic Profiles

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Day(s) Performed

Monday through Saturday

Method Name

Electrochemiluminescence Immunoassay (ECLIA)

CPT Code Information

86705

Method Description

The Elecsys Anti-HBc (hepatitis B virus core antibody) IgM assay is based on the sandwich immunoassay principle and performed with an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. Anti-HBc IgM present in patient's sample is pretreated with anti-Fdy reagent to block specific IgG. After addition of biotinylated monoclonal human IgM-specific antibodies, the complexes formed from reaction of ruthenium-labeled HBc antigen, streptavidin-coated microparticles (solid phase), anti-HBc IgM present in the sample, and the biotinylated anti-human IgM bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode, and unbound substances are washed away. Voltage is applied to the electrode that induces chemiluminescent emissions, which are measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the sample to the cutoff index value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HBc IgM. Roche Diagnostics; v2.0, 12/2024)

Forms

If not ordering electronically, complete, print, and send 1 of the following:

-Gastroenterology and Hepatology Test Request (T728)

-Infectious Disease Serology Test Request (T916)