Order Code LABAIVC Aldosterone, Inferior Vena Cava, Serum
Additional Codes
Mayo Tst Code: AIVC
Reporting Name
Aldosterone, IVCPerforming Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Patient Preparation: Spironolactone (Aldactone) should be discontinued for 4 to 6 weeks before testing. The plasma renin activity cannot be interpreted if the patient is being treated with spironolactone.
Collection Container/Tube:
Preferred: Red top
Acceptable: Serum gel
Submission Container/Tube: Plastic vial
Specimen Volume: 1.8 mL
Collection Instructions:
1. See Renin-Aldosterone Studies for detailed instructions.
2. The recommended collection time is 8 a.m., after the patient is active for approximately 2 hours. Try to collect the specimen as close to that time as possible and no later than 10 a.m.
3. Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
1 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | |
Frozen | 30 days | ||
Ambient | 4 days |
Special Instructions
Reference Values
No established reference values.
Day(s) Performed
Monday through Friday
CPT Code Information
82088
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Useful For
Investigation using inferior vena cava specimen for:
-Primary aldosteronism (eg, adrenal adenoma/carcinoma and adrenal cortical hyperplasia)
-Secondary aldosteronism (renovascular disease, salt depletion, potassium loading, cardiac failure with ascites, pregnancy, Bartter syndrome)
Method Description
Aldosterone-d6 is added to serum and plasma samples as an internal standard. Aldosterone and aldosterone-d6 are extracted from the specimens using a Strata X cartridge. The eluate is dried down under nitrogen, reconstituted with 70/30 methanol/water containing estriol and analyzed by liquid chromatography tandem mass spectrometry using multiple reaction monitoring in the negative mode.(Fredline VF, Taylor PJ, Dodds HM, Johnson AG: A reference method for the analysis of aldosterone in blood by high-performance liquid chromatography-atmospheric pressure chemical ionization-tandem mass spectrometry. Anal Biochem. 1997 Oct 15;252(2):308-313)