Order Code LABDULOX Duloxetine, Serum
Additional Codes
Mayo Test Code: DULOX
Reporting Name
Duloxetine, SUseful For
Monitoring duloxetine serum concentration during therapy
Evaluating potential duloxetine toxicity
Evaluating patient compliance
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Serum RedSpecimen Required
Collection Container/Tube: Red top (serum gel/SST are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Draw blood immediately before the next scheduled dose (trough).
2. Centrifuge and aliquot serum into plastic vial within 2 hours of collection.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Refrigerated (preferred) | 28 days | |
Ambient | 28 days | ||
Frozen | 28 days |
Reference Values
30-120 ng/mLDay(s) Performed
Wednesday
CPT Code Information
80299
Method Description
Serum samples containing duloxetine are diluted in an aqueous solution containing deuterated internal standard and then injected onto a high-turbulence liquid chromatography system for online extraction. Detection is by tandem mass spectrometry.(Unpublished Mayo method)
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Forms
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.