Order Code LABFL Fluoride, Plasma
Additional Codes
Mayo Test Code: FL
Reporting Name
Fluoride, PUseful For
Assessing accidental fluoride ingestion
Monitoring patients receiving sodium fluoride for bone disease or patients receiving voriconazole therapy
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Plasma HeparinSpecimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Green top (sodium heparin)
Submission Container/Tube: Plastic vial
Specimen Volume: 3 mL
Collection Instructions: Centrifuge and aliquot plasma into a plastic vial. Glass tubes are not acceptable.
Specimen Minimum Volume
1.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Heparin | Refrigerated (preferred) | 14 days | |
Ambient | 14 days | ||
Frozen | 14 days |
Reference Values
<4.1 mcmol/L
Day(s) Performed
Tuesday
CPT Code Information
82735
Method Description
The fluoride electrode consists of a single-crystal lanthanum fluoride membrane and an internal reference bonded into an epoxy body. The crystal is an ionic conductor in which only fluoride ions are mobile. When the membrane is in contact with a fluoride solution, potential develops across the membrane. This potential depends on the level of free fluoride ions in solution and is measured against an external constant reference potential with a digital pH/mV meter.(Instruction manual: Fluoride Ion Selective Electrode User Guide. Thermo Scientific; 09/2016)
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Ion-Selective Electrode (ISE)