Sign in →

Order Code LABMUMGM Mumps Virus Antibody, IgM and IgG, Serum

Additional Codes

Mayo Test ID:  MMPGM

Reporting Name

Mumps Ab, IgM and IgG, S

Useful For

Diagnosis of mumps virus infection

 

Determination of postimmunization immune response of individuals to the mumps vaccine

 

Documentation of previous infection with mumps virus in an individual with no previous record of immunization to mumps virus

Profile Information

Test ID Reporting Name Available Separately Always Performed
MMPM Mumps Ab, IgM, S Yes Yes
MPPG Mumps Ab, IgG, S Yes Yes

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into plastic vial.


Specimen Minimum Volume

0.9 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 14 days
  Frozen  14 days

Reference Values

IgM:

Negative: Index value 0.00-0.79

Reference value applies to all ages.

 

IgG:

Vaccinated: Positive (≥1.1 AI)

Unvaccinated: Negative (≤0.8 AI)

Reference values apply to all ages.

Day(s) Performed

MMPM: Tuesday

MPPG: Monday through Saturday

CPT Code Information

86735-Mumps, IgG

86735-Mumps, IgM

Method Description

IgM:

The SeraQuest mumps IgM assay is an enzyme capture method. Diluted samples are incubated in wells coated with antihuman-IgM monoclonal antibodies. If present, IgM antibodies are captured in the wells. Wells are washed, removing excess sample. Conjugate-antigen complex (mumps antigen in complex with monoclonal antibodies conjugated to horseradish peroxidase) is added, and the wells are incubated. IgM antibodies specific for the antigen will bind the conjugate. Wells are washed, removing excess conjugate. Peroxidase substrate is added, and the wells are incubated. Stop solution is added converting the substrate to a yellow end product, which is read photometrically.(Package insert: Mumps IgM. Quest International; V 04/2018)

 

IgG:

The BioPlex 2200 Mumps IgG assay uses multiplex flow immunoassay technology. Briefly, serum samples are mixed and incubated at 37° C with sample diluent and dyed beads coated with mumps antigen. After a wash cycle, antihuman-IgG antibody conjugated to phycoerythrin (PE) is added to the mixture and incubated at 37° C. Excess conjugate is removed in another wash cycle, and the beads are resuspended in wash buffer. The bead mixture then passes through a detector that identifies the bead based on dye fluorescence and determines the amount of antibody captured by the antigen by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity. Three additional dyed beads, an internal standard bead, a serum verification bead, and a reagent blank bead are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel, and the absence of significant nonspecific binding in serum.(Package insert: BioPlex 2200 System MMRV IgG. Bio-Rad Laboratories; 02/2019)

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Heat-inactivated specimen Reject

Method Name

MMPM: Enzyme Immunoassay (EIA)

MPPG: Multiplex Flow Immunoassay

Forms

If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.