Order Code LABRBIGG Rubella Antibodies, IgG, Serum
Reporting Name
Rubella Ab, IgG, SUseful For
Determining immune status to the rubella virus
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Instructions: Plastic vial
Specimen Volume: 0.5 Ml
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.4 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 14 days | |
Frozen | 14 days |
Reference Values
Vaccinated: positive (≥1.0 AI)
Unvaccinated: negative (≤0.7 AI)
Reference values apply to all ages.
Day(s) Performed
Monday through Saturday
CPT Code Information
86762
Method Description
The BioPlex 2200 Rubella IgG assay uses multiplex flow immunoassay technology. Briefly, serum samples are mixed and incubated at 37° C with sample diluent and dyed beads coated with rubella antigen. After a wash cycle, anti-human IgG antibody conjugated to phycoerythrin (PE) is added to the mixture and incubated at 37° C. Excess conjugate is removed in another wash cycle and the beads are resuspended in wash buffer. The bead mixture then passes through a detector which identifies the bead based on dye fluorescence and determines the amount of antibody captured by the antigen based on the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity.
Three additional dyed beads, an internal standard bead, a serum verification bead, and a reagent blank bead are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel and the absence of significant nonspecific binding in serum.(Package insert: BioPlex 2200 System MMRV IgG. Bio-Rad Laboratories; 02/2019)
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Heat-inactivated specimen | Reject |
Method Name
Multiplex Flow Immunoassay (MFI)
Forms
If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.