Order Code LABROGM Measles (Rubeola) Virus Antibody, IgM and IgG, Serum

Additional Codes

EPIC Test Code: LABROGM

Mayo Test Code: ROGM

Reporting Name

Measles (Rubeola) Ab, IgM and IgG,S

Useful For

Diagnosing measles virus infection

 

Determination of immune status of individuals to the measles virus using IgG antibody testing

 

Documentation of previous infection with measles virus in an individual without a previous record of immunization to measles virus

Profile Information

Test ID Reporting Name Available Separately Always Performed
ROM Measles (Rubeola) Ab, IgM, S Yes Yes
ROPG Measles (Rubeola) Ab, IgG, S Yes Yes

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.6 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 14 days
  Frozen  14 days

Reference Values

IMMUNOGLOBULIN M

Negative

Reference values apply to all ages.

 

IMMUNOGLOBULIN G

Vaccinated: positive (≥1.1 AI)

Unvaccinated: negative (≤0.8 AI)

Reference values apply to all ages.

Day(s) Performed

Monday through Saturday

CPT Code Information

86765-Rubeola IgM

86765-Rubeola IgG

Method Description

Immunoglobulin M:

The presence of IgM-class antibody to measles is determined by an indirect immunofluorescence assay. After removal of IgG by specific immunoglobulin antibody, the serum is incubated with measles antigen, which is adhered to a glass microscope slide. Antibodies, if present, will bind to the antigen forming stable antigen-antibody complexes. If no antibodies are present, the complexes will not be formed, and the serum components will be washed away. Fluorescein-labeled antihuman-IgM antibody is added to the reaction side and binds to IgM antibodies if present. This results in a positive reaction of bright apple-green fluorescence when viewed with a fluorescence microscope.(Package insert: Measles Virus Antigen Substrate Slide. AESKU.BION; 09/2019)

 

Immunoglobin G:

The BioPlex 2200 Measles IgG assay uses multiplex flow immunoassay technology. Briefly, serum samples are mixed and incubated at 37° C with sample diluent and dyed beads coated with measles antigen. After a wash cycle, antihuman-IgG antibody conjugated to phycoerythrin (PE) is added to the mixture and incubated at 37° C. Excess conjugate is removed in another wash cycle and the beads are resuspended in wash buffer. The bead mixture then passes through a detector that identifies the bead based on dye fluorescence and determines the amount of antibody captured by the antigen based on the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity.

 

Three additional dyed beads, an internal standard bead, a serum verification bead, and a reagent blank bead, are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel and the absence of significant nonspecific binding in serum.(Package insert: BioPlex 2200 System MMRV IgG. Bio-Rad Laboratories; 02/2019)

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Heat-inactivated specimen Reject

Method Name

ROM: Immunofluorescence Assay (IFA)

ROPG: Multiplex Flow Immunoassay (MFI)

Forms

If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.