Sign in →

Order Code LABTLCU Immunoglobulin Total Light Chains, Urine

Additional Codes

Mayo Test Code: TLCU
EPIC Order Code: LABTLCU

Reporting Name

Immunoglobulin Total Light Chains,U

Useful For

Monitoring patients whose urine demonstrates large M-spikes

 

Confirming the quantitation of specimens that show M-spikes by electrophoresis

 

Detecting urine monoclonal proteins and identification of specimens that need urine protein electrophoresis

Profile Information

Test ID Reporting Name Available Separately Always Performed
KTLCU Kappa Total Light Chain, U No Yes
LTLCU Lambda Total Light Chain, U No Yes
KLTRU Kappa/Lambda TLC Ratio, U No Yes

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Urine


Ordering Guidance


If serum is being submitted on the same patient for FLCS / Immunoglobulin Free Light Chains, Serum; that test should be ordered under a different order number.



Specimen Required


Submit only 1 of the following specimens:

Supplies:  Sarstedt Aliquot Tube, 5 mL (T914)

 

Specimen Type: Random urine

Collection Container/Tube: Clean, plastic urine collection container

Submission Container/Tube: Plastic, 5-mL tube

Specimen Volume: 1 mL

Collection Instructions:

1. Collect a random urine specimen.

2. Refrigerate after collection and send refrigerate.

 

Specimen Type: 24-Hour urine

Container/Tube: Plastic, 5-mL tube

Specimen Volume: 1 mL

Collection Instructions:

1. Collect urine for 24 hours.

2. No additive needed for preservation.

3. Urine may be kept ambient during the collection period but should be refrigerated within 4 hours of collection completion.

Additional Information: See Urine Preservatives-Collection and Transportation for 24-Hour Urine Specimens for multiple collections.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Urine Refrigerated (preferred) 7 days
  Frozen  20 days
  Ambient  72 hours

Reference Values

KAPPA TOTAL LIGHT CHAIN

<0.9 mg/dL

 

LAMBDA TOTAL LIGHT CHAIN

<0.7 mg/dL

 

KAPPA/LAMBDA RATIO

0.7-6.2

Day(s) Performed

Monday through Friday

CPT Code Information

83883 x 2

Method Description

In this Siemens Nephelometer II method, the light scattered onto the antigen-antibody complexes is measured. The intensity of the measured scattered light is proportional to the amount of antigen-antibody complexes in the sample under certain conditions. If the antibody volume is kept constant, the signal behaves proportionally to the antigen volume.

 

A reference curve is generated by a standard with a known antigen content on which the scattered light signals of the samples can be evaluated and calculated as an antigen concentration. Antigen-antibody complexes are formed when a sample containing antigen and the corresponding antiserum are put into a cuvette. A light beam is generated with an LED, which is transmitted through the cuvette. The light is scattered onto the immuno-complexes that are present. Antigen and antibody are mixed in the initial measurement, but no complex is yet formed. An antigen-antibody complex is formed in the final measurement.

 

The result is calculated by subtracting value of the final measurement from the initial measurement. The distribution of intensity of the scattered light depends on the ratio of the particle size of the antigen-antibody complexes to the radiated wavelength.(Instruction manual: Siemens Nephelometer II. Siemens, Inc; Version 2.3, 2008; Addendum to the Instruction Manual 2.3, 08/2017)

Reject Due To

All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Method Name

KTLCU, LTLCU: Nephelometry

Urine Preservative Collection Options

Note: The application of temperature controls must occur within 4 hours of completion of the collection.

Ambient (No additive)

OK <72 hours

Refrigerate (No additive)

Preferred

Frozen (No additive)

OK

50% Acetic Acid

No

Boric Acid

No

Diazolidinyl Urea

No

6M Hydrochloric Acid

No

6M Nitric Acid

No

Sodium Carbonate

No

Thymol

No

Toluene

No