Sign in →

Order Code MGRNA Neisseria gonorrhoeae, Miscellaneous Sites, Nucleic Acid Amplification, Varies

Reporting Name

N. gonorr, Misc, Amplified RNA

Useful For

Detecting Neisseria gonorrhoeae in non-US Food and Drug Administration-approved specimen types

 

This test is not intended for use in medico-legal applications.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Varies


Ordering Guidance


This test is used for specimens that are not US Food and Drug Administration (FDA) approved for this assay. Acceptable non-FDA-approved specimen types are ocular swabs and peritoneal fluid.

 

For FDA-approved specimen types, order GCRNA / Neisseria gonorrhoeae, Nucleic Acid Amplification, Varies.



Necessary Information


Specimen source is required.



Specimen Required


Submit only 1 of the following specimens:

 

Specimen Type: Ocular (corneal/conjunctiva)

Supplies:

Aptima Unisex Swab Collection Kit (T583)

Aptima Multitest Swab Specimen Collection Kit (T584)

Container/Tube: Aptima Multitest Swab or Aptima Collection Unisex Swab

Specimen Volume: Swab

Collection Instructions:

1. Swab site using Aptima Multitest Swab or Aptima Collection Unisex Swab. Specimens must be collected using either of these options.

Note: The white swab provided within the collection kit is a cleaning swab and should not be used for collection. Discard the white cleaning swab.

2. Place collection swab in transport tube provided in collection kit. Snap off swab at score line so swab fits into closed tube.

3. Cap tube securely and label tube with patient's entire name and collection date and time.

4. Maintain swab container at 2 to 30° C (refrigerate temperature is preferred), transport within 60 days of collection.

 

Specimen Type: Peritoneal fluid (pelvic wash, cul-de-sac fluid)

Supplies: Aptima Thin Prep Transport Tube (T652)

Container/Tube: Aptima specimen transfer tube

Specimen Volume: 1 mL

Collection Instructions:

1. Transfer 1 mL of specimen into the Aptima specimen transfer tube within 24 hours of collection.

2. Cap tube securely and label tube with patient's entire name and collection date and time.

3. Maintain Aptima specimen transfer tube at 2 to 30° C (refrigerate temperature is preferred), transport within 30 days of collection.


Specimen Minimum Volume

See Specimen Required

Specimen Stability Information

Specimen Type Temperature Time Special Container
Varies Refrigerated (preferred) APTIMA VIAL
  Ambient  APTIMA VIAL
  Frozen  APTIMA VIAL

Reference Values

Negative

Day(s) Performed

Monday through Sunday

CPT Code Information

87591

Method Description

The Hologic Aptima Combo 2 Assay combines the technologies of target capture, transcription-mediated amplification, and dual kinetic assay. The detection of the ribosomal RNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent DNA probes are labeled and combined with amplicon to form stable RNA:DNA hybrids. Light emitted from the labeled RNA:DNA hybrids is measured as photon signals in a luminometer.(Package insert: APTIMA Combo 2 Assay, AW-25929-001. Hologic, Inc; Rev. 002, 06/2023)

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Method Name

Transcription Mediated Amplification

Forms