Order Code RIVAR Rivaroxaban, Anti-Xa, Plasma
Ordering Guidance
This assay is not indicated for monitoring low-molecular-weight heparin (LMWH) or unfractionated heparin (UFH) concentrations. The presence of UFH and LMWH will cause the rivaroxaban anti-Xa level to be falsely elevated.
This assay is optimized to measure rivaroxaban concentration in presence of coagulation factor Xa recombinant, inactivated-zhzo (andexanet alfa, Andexxa).
Necessary Information
Specimen Required
Specimen Type: Platelet-poor plasma
Collection Container/Tube: Light-blue top (3.2% sodium citrate)
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Specimen should be collected 2 to 4 hours (peak) after a dose or just prior (trough) to the next dose for rivaroxaban concentrations.
2. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing
3. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.
4. Aliquot plasma into a plastic vial leaving 0.25 mL in the bottom of centrifuged vial.
5. Freeze plasma immediately (no longer than 4 hours after collection) at -20° C or, ideally, below -40 C degrees.
Additional Information:
1. A double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
2. Each coagulation assay requested should have its own vial.
Forms
If not ordering electronically, complete, print, and send a Coagulation Test Request (T753) with the specimen.
Useful For
Measuring rivaroxaban concentration in selected clinical situations (eg, kidney insufficiency, assessment of compliance, periprocedural measurement of drug concentration, suspected overdose, advanced age, and extremes of body weight)
Special Instructions
Method Name
Chromogenic Assay
Reporting Name
Rivaroxaban, Anti-Xa, PSpecimen Type
Plasma Na CitSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Na Cit | Frozen | 42 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Reference Values
An interpretive report will be provided.
Method Description
The rivaroxaban, anti-Xa assay is performed on the Instrumentation Laboratory ACL TOP Family using the HemosIL Liquid Anti-Xa kit. The liquid Anti-Xa kit is a 1-stage chromogenic assay based on a synthetic chromogenic substrate and on factor Xa inactivation. Factor Xa is neutralized directly by rivaroxaban. Residual factor Xa is quantified with a synthetic chromogenic substrate. The para-nitroaniline released is monitored kinetically at 405 nm and is inversely proportional to the rivaroxaban in the sample.(Package insert: HemosIL Liquid Anti-Xa kit. Instrumentation Laboratory Company; REV 06/2017)
Day(s) Performed
Monday through Friday