Order Code SALCT Cortisol, Saliva
Reporting Name
Cortisol, SalivaUseful For
Screening for Cushing syndrome
Diagnosis of Cushing syndrome in patients presenting with symptoms or signs suggestive of the disease
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SalivaNecessary Information
Collection time is required.
Specimen Required
Patient Preparation:
1. Do not brush teeth before collecting specimen.
2. Do not eat or drink for 15 minutes prior to specimen collection.
Supplies: Cortisol, Saliva Collection Kit (T514)
Container/Tube: SARSTEDT Salivette
Specimen Volume: 1.5 mL
Collection Instructions:
1. Provide patient with a Saliva Collection Kit (Salivette) containing the Cortisol - Saliva Collection Instructions and ask them to follow the instructions as written.
2. Instruct patient to collect specimen between 11 p.m. and midnight and record collection time on the Cortisol - Saliva Collection Instructions sheet.
3. Instruct patient to return Cortisol - Saliva Collection Instructions with the appropriately labeled Salivette to the laboratory.
Additional Information:
1. Reference values are also available for an 8 a.m. (7 a.m.-9 a.m.) or a 4 p.m. (3 p.m.-5 p.m.) collection, however, the 11 p.m. to midnight collection is preferred.
2. If multiple specimens are collected, submit each vial under a separate order.
Specimen Minimum Volume
0.6 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Saliva | Refrigerated (preferred) | 28 days | |
Frozen | 60 days | ||
Ambient | 28 days |
Reference Values
7 a.m.-9 a.m.: 100-750 ng/dL
3 p.m.-5 p.m.: <401 ng/dL
11 p.m.-midnight: <100 ng/dL
Day(s) Performed
Monday through Friday
CPT Code Information
82533
Method Description
Isotopically labeled cortisol and cortisone are added to the sample as internal standards. Cortisol, cortisone, and coinciding internal standards are extracted from the specimens using online turbulent flow chromatography, which is followed by liquid chromatography tandem mass spectrometry analysis. Although cortisone is not reported, it is used to verify the cortisol value and screen for possible contamination from exogenous sources.(Unpublished Mayo method)
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Special Instructions
Forms
If not ordering electronically, complete, print, and send General Request (T239)