Sign in →

Order Code SGSU Synthetic Glucocorticoid Screen, Random, Urine

Reporting Name

Synthetic Glucocorticoid Screen, U

Useful For

Confirming the presence of the listed synthetic glucocorticoids

 

Confirming the cause of secondary adrenal insufficiency

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Urine


Ordering Guidance


 



Specimen Required


Supplies: Urine tubes, 10 mL (T068)

Container/Tube: Plastic, 10-mL urine tube

Specimen Volume: 5 mL

Collection Instructions:

1. Collect a random urine specimen.

2. No preservative.


Specimen Minimum Volume

0.6 mL

Specimen Stability Information

Specimen Type Temperature Time
Urine Frozen 14 days

Reference Values

Negative

Cutoff concentrations

Betamethasone: 0.10 mcg/dL

Budesonide: 0.20 mcg/dL

Dexamethasone: 0.10 mcg/dL

Fludrocortisone: 0.10 mcg/dL

Megestrol acetate: 0.10 mcg/dL

Methylprednisolone: 0.10 mcg/dL

Prednisolone: 0.10 mcg/dL

Prednisone: 0.10 mcg/dL

Triamcinolone acetonide: 0.10 mcg/dL

 

Values for normal patients not taking these synthetic glucocorticoids should be less than the cutoff concentration (detection limit).

Day(s) Performed

Wednesday

CPT Code Information

80299

LOINC Code Information

Test ID Test Order Name Order LOINC Value
SGSU Synthetic Glucocorticoid Screen, U 46959-3

 

Result ID Test Result Name Result LOINC Value
23562 Betamethasone 46946-0
23563 Budesonide 46947-8
23564 Dexamethasone 46948-6
23565 Fludrocortisone 46949-4
23569 Megestrol Acetate 46953-6
23570 Methylprednisolone 46954-4
23571 Prednisolone 46955-1
23572 Prednisone 46956-9
23574 Triamcinolone Acetonide 46958-5

Method Description

The synthetic glucocorticoids are extracted from 0.5 mL of urine using an acetonitrile protein precipitation followed by methylene chloride liquid extraction of the solvent. Cortisol-9, 11, 12, 12-d, and triamcinolone-d1 acetonide-d6 are added to each sample before the liquid extraction and serve as the internal standards. Then, 17 mcL of the reconstituted sample extract is injected into a high-performance liquid chromatography system and analyzed by tandem mass spectrometry. The mass spectrometer has an electrospray interface and is operated in the multiple reaction monitoring positive mode. The calibration utilizes a 4-point standard curve over a concentration range of 0 to 25 mcg/dL.(McWhinney BC, Ward G, Hickman PE. Improved HPLC method for simultaneous analysis of cortisol, 11-deoxycortisol, prednisolone, methylprednisolone, and dexamethasone in serum and urine. Clin Chem. 1996;42:979-981; Savu S, Silvestro L, Haag A, Sorgel F. A confirmatory HPLC-MS/MS method for ten synthetic corticosteroids in bovine urines. J Mass Spectrom. 1996;31[12]:1351-1363; Djedovic NK, Rainbow SJ. Detection of synthetic glucocorticoids by liquid chromatography-tandem mass spectrometry in patients being investigated for Cushing's syndrome. Ann Clin Biochem. 2011;48[Pt 6]:542-549. doi:10.1258/acb.2011.010250)

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Method Name

Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)